INTRODUCTION
Spirometry is the standard method to measure the airway obstruction in chronic lung diseases
1, 2). In some situations, however, spirometric measurement of lung function is difficult largely because of inability to perform adequate forced expiration or lack of patient’s cooperation. Whole body plethysmography is a more advanced method, but it also needs the patient’s cooperation and requires expensive devices
3).
In 1956, DuBois et al.
4) first described an oscillatory method for measurement of the respiratory system mechanics, but this technique was not accepted widely because of its technical difficulties and the greater attractiveness of the body plethysmography. Recent advances in microprocessor technology, however, have solved many earlier problems, and Forced oscillation Techniques (FOT) becomes more widely used in clinical pulmonary laboratories
5).
FOT, which characterize the mechanical properties of the respiratory system over a wide range of frequencies, is a non-invasive and effort-independent method for measuring airway resistance. The primary measurements of FOT are the pressure and flow generated at a given frequency, and the instantaneous pressure-flow relationship, namely impedance of the respiratory system (Zrs), is further analyzed into a real part or resistance (Rrs) and an imaginary part or reactance (Xrs)
5, 6).
An important advantage and the clinical potential of FOT is that it is rapid and demands only the patient’s passive cooperation. Hence, FOT is proved to be feasible for lung function measurement in infants
7) and young children
8, 9), for challenge tests
10), for reversibility tests of airflow obstruction, and for large epidemiologic surveys
5).
Treatment of bronchial asthma is based on its severity which is determined by symptoms, spirometric measurement of forced expiratory volume at 1 second (FEV
1) and diurnal variation in peak expiratory flow (PEF)
11). Some patients with asthma may describe obscure symptoms and inadequately perform spirometry, so proper estimation of asthma severity is often impossible. The FOT is known to be more sensitive to lower airway obstruction, such as bronchial asthma, than to interstitial lung disease and chest wall pathology
5).
This study was performed to evaluate the usefulness for the clinical applications of FOT in patients with bronchial asthma by estimating the associations between asthma severity and FOT parameters, and the relationships between FOT and spirometry parameters.
MATERIALS AND METHODS
1. Patients
Of the patients who visited our allergy clinics for the evaluation of respiratory complaints from July 1997 to June 1999, 216 patients with bronchial asthma were enrolled in this study. Asthma was clinically diagnosed on the basis of typical respiratory symptoms, such as dyspnea, wheezing, coughing or presence of nocturnal aggravation plus one of the following Two; 1) reversible airflow limitations-improvement of FEV1 of more than 12% after inhalation of 200 μg of salbutamol MDI or long-term anti-inflammatory treatment, or 2) significant nonspecific airway hyperresponsiveness-methacholine PC20 less than 25 mg/mL. Among 216 patients, male:female ratio was 102:114, mean age was 46.3±13.8 years (age range 19 to 70 years), and 77 patients had concurrent rhinitis.
Baseline data, including age, sex, height and asthma symptoms were obtained from each patient at the time of procedures. Then, patients were classified into mild asthma (mild group, n=111, 51.3%), moderate persistent asthma (moderate group, n=78, 36.1%) and severe persistent asthma (severe group, n=27, 12.6%) according to NHLBI guideline
11) (
Table 1).
2. Forced Oscillation Techniques and Spirometry
FOT measurements were performed with a pneumotachograph and its built-in pressure transducer using commercially available equipment, MS-IOS Digital Instrument (Master Screen IOS version 4.32, Jaeger, Würzburg, Germany). The patient was seated comfortably with the neck slightly extended. Measurements were carried out during quiet breathing through a mouthpiece, and the cheeks were supported by the patient’s palm and finger and the nose was occluded with nose clips to minimize the upper airway artifact. After adaptation to this procedure, FOT were processed for 30 second tidal breathing. The digitized pressure and flow signals were sampled and, after low pass filtering to prevent aliasing, the signals were fed into a fast Fourier transformation analyzer. Finally, FOT parameters, impedance (Zrs), respiratory resistance at 5 Hz, 20 Hz and 35 Hz (R5Hz, R20Hz, R35Hz respectively), reactance at 5 Hz (X5Hz) and resonant frequency (Fres), were calculated and measured from each impulse.
Lung volumes and flows were measured in the sitting position, using the same device, after FOT. For each patient, spirometry was performed 3 times and the best value for FEV1, forced vital capacity (FVC), PEF and MMEF (maximal mid-expiratory flow rate) from three flow-volume curves were selected.
3. Nonspecific Airway Hyperresponsiveness
Nonspecific airway hyperresponsiveness was determined by methacholine challenge with a modified method of Chai et al.
12). Methacholine was diluted to physiologic saline and delivered by dosimeter (DSM-2, S&M Instrument Company Inc., Doylestown, PA, USA). Each patient took 5 full inhalations from functional residual capacity. The initial concentration of methacholine administration was 0.075 mg/mL and a dose response curve was constructed by administering doubling concentrations of methacholine up to 25 mg/mL. Provocational concentrations of methacholine used to produce a 20% decrease in FEV
1 from the baseline value, M-PC
20, was calculated by linear interpolation between the last two points on the dose response curve.
4. Statistical Analysis
The SPSS statistical package 8.0 was used to analyze the data. Pearson’s correlation was used to assess the association between FOT and spirometry parameters. Geometric mean of M-PC20 and serum IgE was calculated from logarithmic transformation of each value. Comparisons of eosinophil counts, serum IgE, M-PC20, spirometry parameters and FOT parameters according to asthma severity group were analyzed by ANOVA. Using Pearson’s correlation, we estimated reproducibility of FOT and spirometry measurements by comparison of the same parameters of two repeated tests. All values are expressed as mean± standard deviation, and p values less than 0.05 were considered significant.
DISCUSSION
Classification of asthma severity is based on the degree of airway hyperresponsiveness, eosinophil counts and/or eosinophil cationic protein levels in blood or sputum, and lung functions. Measurements of lung function are objective and easily available method to assess airway obstruction, and are recommended as useful criteria for the assessment of asthma severity by recent guidelines
11). Spirometry is the standard method for measurements of lung function and its validity, reproducibility and responsiveness is well verified
1). Of the several parameters in spirometric measurements, FEV
1 is the most sensitive and reproducible index to detect airway obstruction
14). MMEF is effort independent and partially reflects small airway obstruction, but its sensitivity and reproducibility is lower than FEV
114). Occasionally, spirometric measurement of lung function is difficult due to lack of patient’s cooperation. Whole body plethysmography is another useful technique to measure airway resistance, but it also needs the patient’s cooperation
3).
FOT is an alternative method to overcome these problems, because it requires only passive cooperation. The basic principle of FOT is measurement of the relationship between pressure waves applied externally to the respiratory system and the resulting respiratory air flow
6). The ratio of the amplitude of the pressure wave to the amplitude of the flow wave constitutes the impedance, and then it is further analyzed into resistance (Rrs) and reactance after Fourier transformation
5, 6).
Respiratory resistance is relatively constant at different frequencies in healthy subjects, but it is increased at the lower applied frequencies but falls with increasing applied frequencies in patients with airway obstruction
13). This resistance fall with increasing frequency is a result of the presence of inhomogeneity of pulmonary mechanical properties, and is an expected and similar finding in both asthma and chronic obstructive lung disease. Values of reactance are negative at lower frequencies, and progressively increase and become positive at higher frequencies. The frequency at which reactance becomes zero is resonant frequency, and it is increased in proportion to the degree of the airway obstruction. So, these parameters are usefully applied to assess the degree of the airway obstruction
5, 15, 16).
In this study, we tried to estimate the clinical usefulness of FOT in the classification of asthma severity. We did not differentiate patients with mild intermittent asthma from mild persistent asthma, because both groups showed the same range of spirometry measurements, and FOT could not classify mild asthma into two distinct groups in our unpublished data. We also included smokers in this study, because FOT difference between smokers and non-smokers was not found in our analysis (data not shown) and in a previous report by Landser et al.
17).
As expected, the degree of the airway hyperresponsiveness and spirometric measurements of lung function showed significant difference according to asthma severity, and these are quite reasonable because we divided the patients into 3 severity groups with mainly FEV
1 criteria. Sputum eosinophil counts, however, were lower in the severe group than in the less severe group. These findings might be the results from treatments of systemic or high doses inhaled corticosteroid and consequent suppression of eosinophils in the severe group
18).
In general, resistance at lower frequency is a very sensitive index to detect airway obstruction
19). In our study, R5Hz, R20Hz and resonant frequency of FOT measurements showed significant difference according to asthma severity as spirometry parameters. Respiratory impedance in the severe group was also higher than in the mild group. Based on the frequency dependency of resistance, we calculated a new index, resistance difference between two different frequencies (resistance at low frequency - resistance at relatively high frequency), to determine the inhomogeneity of mechanical properties of the obstructed lungs. Such an index, R5Hz–R20Hz, also showed significant difference between each group and was considered as a useful parameter of FOT. But it is still a conceptual meaning, and further studies will be needed to clearly define its value and significance.
Several studies compared FOT with spirometry and/or whole body plethysmography in the assessment of airflow obstruction. Airway resistance measured by FOT (Rrs) and body plethysmography (sRaw) showed good correlation, but relationships between FOT and FEV
1 or MMEF were variable according to each study design
5, 8, 13, 20). In this study, FOT parameters had significant association with FEV
1, and with MMEF in patients with normal pulmonary function. Considering correlation coefficient, however, their association was not strong, and such result was not surprising because each parameter measured its own peculiar part of the respiratory mechanics. We also estimated reproducibility of FOT by comparing the repeated tests performed at 6 months interval in patients with the same clinical condition. The correlation coefficient of R5Hz was 0.78 and very similar with that of FEV
1 (r=0.79). Therefore we can consider that resistance measured by FOT, especially at low frequency, is a highly reproducible index and compatible with FEV
1.
In spite of some attractive advantages of FOT, it has still several limitations in clinical practice
5). The first is that measurements of FOT are influenced by the extrathoracic upper airway artifact. That is, the pressure-flow properties of the upper airway may play a role in determining the total respiratory impedance. To overcome this problem, we advised patients to support their cheeks with their palms, but these methods cannot eliminate upper airway artifact totally. Recently, Farre et al.
21) suggested that change of FOT admittance, not standard generator but head generator, can minimize artifact caused by the upper airway. So it is expected that further development and progress of FOT will take off such an artifact. The second limitation is lack of reference values of each FOT parameter for Korean subjects, because it is being used rather lately in clinical practice, and the number of reports of normal values for oriental people is limited. The third limitation is that it is not effectively sensitive to detect the pathologic conditions of chest wall and restrictive lung diseases. Moreover, the significance of parameters measured at higher frequencies is not clear yet. Finally, some parameters have still not been employed widely because of their obscure meaning. For example, FOT can divide respiratory resistance into two compartments, central resistance and peripheral resistance, but fractionating criteria is not an anatomical structure but a theoretical conception. Therefore such parameters have to be explained clearly based on scientific researches before clinical practice.
In conclusion, FOT is a non-invasive, effort-independent, and easily applicable method to detect airway obstruction in bronchial asthma, and is a useful and alternative method for evaluating the clinical status of patients with asthma. Because asthma is a heterogeneous disease characterized with various clinical and laboratory features, a physician cannot understand a patient’s condition exactly by only one criterion. In this situation, measurements of FOT can be added to other conventional criteria. Further investigations may expand FOT applications in various clinical conditions and diseases.