Cho, Jeong, Cho, Ahn, Cha, Seo, Park, Jeong, Park, and Kang: Atrial Fibrillation in Patients with Permanent VVI Pacemakers: Risk Factors for Atrial Fibrillation

Abstract

Objectives

Atrial fibrillation (AF) does not only deteriorate the cardiac function and increases the thromboembolic risk but also triggers rapid and irregular ventricular rhythm in patients with atrial synchronous pacing. However, the risk factors for the development of AF in patients with pacemakers are not clearly determined yet. The present study was designed to determine the risk factors for AF in patients with VVI pacemakers.

Methods

This study included 80 patients (41 sick sinus syndrome, 39 AV block) who were followed for more than 6 months or developed AF regardless of the duration of follow-up after implantation of VVI pacemakers. Patients were divided into two groups according to whether or not AF developed during follow-up (mean: 25.7±2.5 months): group A developed AF and group B did not. The underlying arrhythmias, cardiovascular risk factors, left atrial size, characteristics of P wave were compared between the two groups.

Results

The mean age of the patients was 58.9±11.4 years and 28(35%) were male. AF developed in 13(16.3%) of 80 patients with VVI pacemakers. Sick sinus syndrome (SSS) as an underlying arrhythmia was significantly more frequent in group A than group B(84.6% vs. 44.8%, p<0.01). P wave width was greater in group A (127.6±24.8ms) than in group B (110.7±17ms) (p<0.05). There was, however, no significant difference in cardiovascular risk factors, left atrial size, P wave axis and amplitude between the two groups.

Conclusion

These results suggest that sinus node dysfunction and intra-atrial conduction delay may be the risk factors for AF in patients with VVI pacemakers. Further studies are needed to determine how sick sinus syndrome and intra-atrial conduction delay increase the risk for AF in patients with VVI pacemakers.

INTRODUCTION

Cardiac pacing therapy has been shown to prevent sudden cardiac death and improve the quality of life in patients with symptomatic bradyarrhythmias since the artificial pacemaker was first introduced in clinical practice in 1958. Cardiac pacing therapy is being widely used as the only definitive mode of therapy to treat symptomatic or significant bradyarrhythmia, such as sick sinus syndrome (SSS) and AV block (AVB)1). As a result of recent advances in pacemaker technology, the function and size of the artificial pacemaker is much improved. Now, highly sophisticated physiologic pacemakers capable of dual chamber pacing and/or rate-responsive pacing are available.
Atrial fibrillation (AF) is one of the major problems developed in patients with permanent cardiac pacing therapy2). AF does not only depress cardiac function with increasing thromboembolic risk but also makes it mandatory to change the pacing mode in patients with atrial tracking ventricular pacing mode, such as DDD and VDD. It is, therefore, very important to identify risk factors for AF to develop preventive strategies3). Pacing mode has been reported as a major risk factor for AF in patients with permanent pacemakers, based on the observation that AF developed more frequently in patients with VVI pacing mode than in patients with DDD pacing mode4). However, the risk factors for AF in patients with permanent pacemakers have not been clearly identified yet. The purpose of this study is to determine the risk factors for AF in patients with VVI pacing.

MATERIALS AND METHODS

One hundred and fifty three patients were implanted with permanent VVI pacemakers at the Chonnam University Hospital from August, 1983 through July, 1993. All pacemakers were implanted transvenously via the right or left subclavian vein. Among them, 80 patients (52.3%) who were followed for more than 6 months or who had complications with AF within 6 months after the implantations of a VVI pacemaker were enrolled in this study. All 80 patients had no prior history of AF. The eighty patients were divided into two groups according to whether or not AF developed during follow-up; AF developed in 13 patients (group A) but did not develop in the remaining 67 patients (group B).
The diagnosis of AF was made by analyzing standard 12-lead electrocardiogram recorded at each follow-up visit. If atrial rhythm was difficult to evaluate, we analyzed atrial rhythm after slowing the pacing rate of the pacemaker down to 40–50/min temporarily using the programmer. In order to investigate the risk factors for AF, underlying cardiac rhythm and cardiovascular risk factors, such as smoking, hypertension, hyperlipidemia and diabetes mellitus, were evaluated and the left atrial size and characteristics of P wave, including mean electrical axis, amplitude and duration, were measured. Left atrial size was measured as an antero-posterior diameter of the left atrium in the M-mode echocardiogram recorded at the left parasternal border under the guidance of 2-dimensional echocardiography. The maximum amplitude and duration of P wave were taken as the values of P wave amplitude and duration and P wave axis was considered abnormal if the axis was not within 0–90 degrees.
Statistical analysis was done using Student's t-test or Chi-square test and p<0.05 was considered significant.

RESULTS

Mean age of the 80 patients was 58.9±11.4 years, 28 (35%) were male. Mean follow-up period was 25.7±2.5 (range: 1–99) months (Table 1). Indications for permanent pacemaker implantation were sick sinus syndrome (SSS) in 41(51.3%), second-degree AV block (AVB) in 5(6.3%), and third-degree AVB in 34(42.5%) patients. Thirteen (16.3%) patients developed AF during the follow-up period.
Eleven (84.6%) of 13 patients in group A and 30 (44.8%) of 67 patients in group B were sick sinus syndrome, showing that AF developed significantly more frequently in patients with SSS than AVB (p<0.01, Table 2). Left atrial size measured using M-mode echocardiography was 40.1±3.6 mm in group A and 36.6±5.7 mm in group B, showing no significant difference between the 2 groups (Table 2). Number and distribution of the cardiovascular risk factors were similar in both groups (1.4±0.3 in group A vs. 1.5±0.5 in group B) (Table 2). In 12-lead electrocardiogram, recorded before pacemaker implantation, the incidence of abnormal P wave axis and P wave amplitude were similar in the 2 groups. However, P wave duration was greater in group A than group B (127.6±24.8 ms vs. 110.7±17.0 ms, p<0.05) (Table 2).

DISCUSSION

Cardiac pacing therapy is a mainstay in the treatment of symptomatic or severe bradyarrhythmia due to sinus node dysfunction or AV conduction disturbance. It has been shown that the permanent pacemaker significantly decreases the morbidity and mortality of SSS and AVB1, 5). The implantation of a permanent cardiac pacemaker is widely performed in many secondary or tertiary hospitals since the first permanent cardiac pacemaker was implanted in Korea in 1968412). In recent days, the improvement of technology makes available highly sophisticated cardiac pacemakers that automatically regulate cardiac pacing according to metabolic demands of patients as the normal sinus node do.
AF, an irregular and very rapid atrial tachyarrhythmia, evokes hemodynamic and thromboembolic complications due to rapid and irregular ventricular rhythm and a loss of atrial mechanical contraction, in addition, AF in patients with atrial tracking dual chamber pacing mode such as DDD or VDD may provoke serious problems by triggering irregular and rapid ventricular pacing. Those pacing modes with atrial tracking function should be changed into VVI if AF develops de novo after implantation of pacemakers, and dual chamber pacing with atrial tracking function should be avoided in patients with persistent AF. Thus, AF is a very important arrhythmia in candidates for pacemaker implantation as well as in patients with pacemakers, and determination of risk factors for AF is very useful to develop preventive measures and to guide pacing therapy3). Previous studies have suggested that the development of AF in patients with cardiac pacemaker is mainly influenced by underlying arrhythmia1417).
This study was performed to determine risk factors for AF in patients with VVI pacemakers for SSS or significant AVB. Our study suggests that SSS as an underlying arrhythmia and prolongation of P wave are risk factors for AF.
In our study, AF developed de novo in 13(16.3 %) of 80 patients with VVI permanent pacemakers during the follow-up period. The incidence of AF in SSS was 26.8% and 5.6% in AVB. This significantly higher incidence of AF in SSS was consistent with the previous studies. Grimm et al.18) reported that AF was more frequent in patients with SSS than in patients with AVB. The development of AF, in patients with VVI pacemakers, is thought to be related to excessive loading of the atria, due to dissociation between atrial and ventricular contractions, or atrial contraction against the close AV valves, due to retrograde conduction of paced ventricular impulses to the atria, and development or progression of atrial disease itself4, 19, 20). Therefore, the higher incidence of AF in SSS than in AVB may reflect that SSS is more frequently associated with atrial disease and AVB with retrograde ventriculoatrial conduction block. Several studies have shown that DDD or AAI pacing has less chance to have AF during the follow-up than VVI pacing2123). DDD or AAI pacing is preferable to VVI pacing, particularly in high risk patients for AF, for lowering the incidence of AF after pacemaker implantation.
This study demonstrated that intra-atrial conduction delay is also a risk factor for AF. Snoeck et al.21) reported that the future occurrence of AF could be predicted by P wave recorded before pacemaker implantation, and AF was hardly induced in patients with the width of the widest P wave less than 100 msec. In our study, the width of P wave in patients with AF was significantly wider than that in patients without AF (127.6±24.8 ms vs. 110.7±17.0ms). But, the electrical axis and the amplitude of P wave were similar whether AF developed or not. Our observation that P wave duration, reflecting intra-atrial conduction time, is a risk factor for AF is not unusual in that AF is induced and maintained by 4–6 reentrant wavelets in the atria and conduction delay, in addition to undirectional block, is a prerequisite for reentry. Recently, signal-averaged P wave duration, which is measured by the signal-averaging technique of P wave, has been proposed as a rapid and noninvasive marker of AF24). Atrial dilatation is known to be a risk factor for AF. So, we examined the left atrial size using M-mode echocardiography under the guidance of 2-dimensional echocardiography. However, contrary to our expectation, left atrial size was not significantly different between patients with and without AF. This finding suggests that atrial size does not influence atrial vulnerability to AF until it reaches a certain degree.
In conclusion, the risk factors for atrial fibrillation in patients with VVI pacemakers are sick sinus syndrome and intra-atrial conduction delay. Further studies are needed to determine how sick sinus syndrome and intra-atrial conduction delay increase the risk for AF in patients with VVI pacemakers.

Table 1.
Baseline Characteristics of the 80 Patients with VVI Pacemakers
Mean age, years 58.9 ± 11.4
Male/Female 28/52 (1:1.9)
Underlying arrhythmias
 Sick sinus syndrome 41 (51.3%)
 AV block 39 (48.8%)
  - 2nd degree 5 (6.3%)
  - 3rd degree 34 (42.5%)
Follow-up duration, months 25.7 ±2.5
Table 2.
Comparison of Clinical Characteristics Between the Patients Complicated with AF (group A) and no AF (group B) During the Follow-up Period
Group A (n=13) Group B (n=67) P value
Age, years 55.1±4.3 59.6 ± 11.9 NS
Sex (M/F) 3/10 (1:3.3) 25/42 (1:1.7) NS
Indication (%) <0.01
 Sick sinus syndrome 11 (84.6) 30 (44.8)
 AV block 2 (15.4) 37 (55.2)
Follow-up, months 26.4 ±6.0 25.6 ±2.8 NS
No of CV risk factors 1.4 ±0.3 1.5±0.5 NS
Left atrial size, mm 40.1 ±3.6 36.6 ±5.7 NS
Characteristics of P waves
 Abnormal axis (%) 2 (15.4) 6 (9.0) NS
 Amplitude, mV 0.22 ±0.06 0.23 ±0.10 NS
 Duration, ms 127.6 ±24.8 110.7 ± 17.0 <0.05

M: male, F: female, CV: cardiovascular, No: number, NS: non-significant

REFERENCES

1. Barold SS. The fourth decade of cardiac pacing hemodynamic, electrophysiological, and clinical con siderations in the selection of, the optimal pace maker. In: Zipes DP, Jalife J, eds. Cardiac Electrophysiology-From Cell to Bedside-2nd ed. 1366–1392Philadelphia: WB Saunders Co, 1995.

2. Cho EK, Choi ES, Park HJ, Kim YT, Kim YJ, Lee SM, Lee HS, Kim BY, Yoo SW. Complications for implanted permanent pacemaker. Korean J of Intern Med 1993;45:299–306.

3. Camm AJ, Katritsis D. Ventricular pacing for sick sinus syndrome - a risky business? PACE 1990;13:695–699.
crossref pmid
4. Feuer J, Shandling AH, Messenger JC. Influence of cardiac pacing mode on the long-term deveio pment of atrial fibrillation. Am J Cardiol 64:1989;1376–1379.
crossref pmid
5. Imich W, Parsonnet V, Myers GH. Introduction and historical review. PACE 1978;1:371–374.
crossref pmid
6. Lee DY, Choi HC, Lee KO, Suh SK, Yoo HS. An application of internal artificial pacemaker to com plete A-V block with Adams-Stokes syndrome: re port of a case. Korean J of Intern Med 1969;12:873–876.

7. Lee DY, Yoon YJ, Cho BK, Hong SR. Experience in treating complication after implantation of tran svenous pacemaker. Korean J of Chest Surgery 1976;9:265–270.

8. Park IS, Kim KA, Hong SJ, Kim SS, Kim HJ, Lee HK. Permanent transvenous cardiac pacemaker implantation: clinical observation in 14 cases of various cardiac conduction defects. Korean J of intern Med 1979;22:457–265.

9. Kwack MS, Na BW, Kim SS, Lee HK. Pervenous intracardiac pacemaker implantation. J of Korean Med Assoc 1982;25:359–372.

10. Yoon SH, Shin DK, Kim MC, Park SH, Kim JJ, Kim MJ, Kim KY, Chu DB. Five cases of sick sinus syndrome and implantation of pacemaker. Korean Circulation J 1983;13:435–442.
crossref
11. Cho MC, Nam HS, Oh BH, Park YB, Choi YS, Lee YW. Measurement of pacing threshold of the permanent pacemaker by Vario system. Korean J of Intern Med 1986;31:740–748.

12. Cho JG, Jeong MH, Park JC, Yang SJ, Park CH, Gill GC, Cho KW, Kang JC. Clinical experience with permanent pacemaker implantations. Korean J of Intern Med 1989;36:1–7.

13. Marthon JW, Eisenhauer AC, Steinhaus DM. Cardiac pacing. In: Eagle KA, Haber E, Desanctis RW, Austen WG, eds. The Practice of Cardiology. 2nd ed. 287–336Boston: Little, Brown and Company, 1989.

14. Kannel WB, Abbott RD, Savage DD, McNamara PM. Epidemiologic features of chronic atria fibrillation. N Engl J Med 1982;306:1018–1022.
crossref pmid
15. Snoeck J, Decoster H, Verherstraeten M, Berkhof M, Vorlat A, Heuten H, Vrints C, Kahn JC, Marchand X. Evolution of P wave characteristics afte pacemaker implantation. PACE 1990;13:2091–2095.
crossref pmid
16. Hesselson AB, Parsonnet BV, Bernstein AD, Bon-avita GJ. Deleterious effects of long-term single-chamber ventricular pacing in patients with sick sinus syndrome: the hidden benefits of dualcha mber pacing. J Am Coll Cardiol 1992;19:1542–1549.
crossref pmid
17. Lamas GA, Estes III NM, Schneller S, Flaker GC. Does dual chamber or atrial pacing prevent atria fibrillation? The need for a randomized controlled trial. PACE 1992;15:1109–1113.
crossref pmid
18. Grimm W, Langenfeld H, Maisch B. Symptoms, cardiovascular risk profile and spontaneous ECG in paced patients: a five-year follow-up study. PACE 1990;13:2086–2090.
crossref pmid
19. Mitsuoka T, Kenny RA, Yeung TA, Chan SL, Perrins JE, Sutton R. Benefits of dual chamber pacing in sick sinus syndrome. Br Heart J 1988;60:338–347.
crossref pmid pmc
20. Boon NA, Frew AJ, Johnston JA, Cobbe SM. A comparison of symptoms and intra-arterial ambu latory blood pressure during long term dual chambe atrioventricular synchronous (DDD) and ventricula demand (VVI) pacing. Br Heart J 1987;58:34–39.
crossref pmid pmc
21. Snoeck J, Decoster H, Vrints C, Marchand X, Kahn JC, Verherstraeten M, Berkhof M, Franque L. Predictive value of the P wave at implantation for atria fibrillation after VVI pacemaker implantation. PACE 1992;15:2077–2083.
crossref pmid
22. Rosenqvist M, Brandt J, Schuller H. Atrial versus ventricular pacing in sinus node disease: A treat ment comparison study. Am Heart J 1986;111:292–297.
crossref pmid
23. Zanini R, Facchinetti AI, Gallo R, Cazzamalli L, Bonandi L. Morbidity and mortality of patients with sinus node disease: Comparative effects of atria and ventricular pacing. PACE 1990;13:2076–2079.
crossref pmid
24. Guidera SA, Steinberg JS. The signal-averaged P wave duration: a rapid and noninvasive marker o risk of atrial fibrillation. J Am Coll Cardiol 1993;21:1645–1651.
crossref pmid
TOOLS
METRICS Graph View
  • 2 Crossref
  • 4 Scopus
  • 8,244 View
  • 61 Download
Related articles

Editorial Office
101-2501, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-2271-6792    Fax: +82-2-790-0993    E-mail: kaim@kams.or.kr                

Copyright © 2024 by Korean Association of Internal Medicine.

Close layer
prev next